Total wrist arthroplasty with the Freedom prothesis: A short-term follow-up.

No Thumbnail Available
Authors
Brown, Matthew
Wharton, Rupert
Begum, Rumina
Trail, Ian
Hayton, Mike
Talwalkar, Sumedh
Issue Date
2024-01-15
Keyword
Type
Article
Citation
J Hand Surg Eur vol 49(1):40-47
PubMed ID
Other Identifiers
Abstract
We retrospectively reviewed our series of primary total wrist arthroplasty with the Freedom® prosthesis. The primary outcome measure was revision, and secondary measures included radiographic loosening, pain, complications, movement range, grip strength and patient-reported measures. We reviewed 12 implants in 11 patients (mean age 59 years, range 45-80) with a mean radiological and clinical follow-up of 2.7 and 3 years, respectively. One radial component failed to integrate and was revised at Day 84. Four carpal components demonstrated areas of lucency. There was a statistically significant reduction in pain, and total flexion-extension increased. Despite high patient satisfaction on a ten-point visual analogue scale score (mean 8.7 out of 10), the mean patient-rated wrist evaluation, Quick Disabilities of the Arm, Shoulder and Hand and Patient Evaluation Measure scores were 52, 55 and 53, respectively. The Freedom® implant reduced pain and preserved wrist movement in our patients; however, annual surveillance is recommended due to the high incidence of early carpal component lucency.Level of evidence: IV.